Bedside Setup Wizard
STEP 1 OF 4
Choose patient
🧑
Adult
Standard adult ranges
🧒
Pediatric
Weight/age-adjusted rates
Choose pacing type
🩸
Temporary Transvenous
Central venous lead to endocardium
❤️
Epicardial
Post-surgical wires (e.g. post-CABG)
Choose chamber(s)
🫀
Ventricular
Single chamber — V lead only
🫀
Atrial
Single chamber — A lead only
🫀🫀
Dual Chamber
A + V leads, AV synchrony possible
Choose underlying rhythm / indication
Device Panel
LIVE SIMULATOR
Capture: —
Sensing: —
Mode: VVI
Guided Capture Threshold Test
STEP-BY-STEP
1Set rate ≥10 ppm above intrinsic rate so continuous pacing is forced.
Current test rate: 90 ppm —
2Slowly decrease output and check for capture after each change.
Output: 10.0 mA
No readings yet.
Guided Sensing Threshold Test
STEP-BY-STEP
1Set rate ≥10 ppm below intrinsic and
output to 0.1 mA to reveal intrinsic rhythm safely.
Test rate: 60 ppm (output fixed at 0.1 mA for this test)
2Slowly increase sensitivity value (mV) — device gets
less sensitive — until the SENSE light stops flashing.
Sensitivity: 1.0 mV
No readings yet.
ECG Behavior Explorer
4 SCENARIOS
Select a scenario to see how the tracing changes — this uses the live device settings from the Device Panel.
Normal Capture
Every pacing spike is immediately followed by a QRS complex — the myocardium is depolarizing in response to each stimulus.
Interactive Troubleshooting
DECISION TREE
What are you seeing on the monitor?
Disease / Indication Library
TAP TO LOAD PRESET